GNP334-6/3P Specifications

Geiger-Mueller Tube
Platinum Plated High Sensitivity,
High Temperature Radiation Couter Tube

Electrical Characteristics

Recommended Operating Voltage (volts)900
Operating Voltage Range (volts)850 - 1000
Maximum Dead Time (μs)150
Maximum Plateau Slope (%/100 volts)5
Minimum Anode Resistor (MΩ)3.3
Recommended Anode Resistor (MΩ)5.0
Gamma Sensitivity Cs137 (cps/mR/hr)110
Max. Background Shielded 50mmPb+3mmAl (cpm)25
Tube Capacitance (pF)8

Mechanical Characteristics

Nominal Active Length (inch/mm)5.00 / 127.0
Window Thickness (mg/cm2)-
Active Diameter (inch/mm)0.73 / 18.5
Cathode MaterialCr - Fe
Cathode Wall Thickness (inch/mm)0.010 - 0.25
Operating Temperature Range (°C)-40 to 150
Fill GasNe + Halogen
Type of Connector3-Pin
Approx. Weight (grams)34
Maximum Altitude (ft)TBD

Equivalent models: NP334-6/3P

Global Nucleonics has the design and engineering capability to produce detectors to your specifications.